Zero Tare (Resets the zero flow rate of the meter)
-
With the meter in manual mode, drive the valve completely closed.
-
Enter the setup menu and enter P6.10
-
Change the value to 1 and press Select to save.
-
-
Max Tare (Sets the max position the valve can drive to in valve position mode)
-
With the meter in manual mode, drive the valve open to the desired max flow rate.
-
Enter the setup menu and enter P6.50
-
Change the value to 1 and press Select to save.
-
-
Altitude Compensation
-
Enter the setup menu and enter P6.60
-
Enter the actual altitude for the installed location.
-
Press Select to save.
-
-
Zero Tare (Resets the zero flow rate of the meter)
-
With the meter in manual mode and supply gas/liquid pressure applied to the meter, drive the valve completely closed.
-
From the main screen, enter the setup menu and push the zero tare button.
-
This will zero the meter and the flow should adjust to zero accordingly.
-
-
Max Tare (Sets the max position the valve can drive to in valve position mode)
-
With the meter in manual mode and supply gas/liquid pressure applied to the meter, drive the valve open to the desired max flow rate.
-
From the main screen, enter the setup menu and push the max tare button.
-
This will set the current position of the valve to be the maximum position allowed during normal operation.
-
-
Altitude Compensation
-
From the main screen, enter the Setup menu and Instrument Configuration - Additional Options menu .
-
Enter the actual altitude for the installed location.
-
Press Select to save.
-
Appendix 3: Startup Procedures
For optimal accuracy and control, the following procedure is recommended:
LED model:
Touchscreen model: